4.6 Connecting the eluent bottle
■■■■■■■■■■■■■■■■■■■■■■
22
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
4.6 Connecting the eluent bottle
The eluent is aspirated out of the eluent bottle via the eluent aspiration
tubing . The eluent aspiration tubing is installed on the entry to the high-
pressure pump.
The tubing must be fed out of the instrument through a suitable opening
(see "Openings on the back panel", page 13) before the loose end can be
connected to the eluent bottle.
Accessories For this step, you need the following accessories:
These parts are part of the Vario/Flex ONE accessory kit (6.5000.010).
■ Eluent bottle (6.1608.070)
■ The eluent bottle cap GL 45 accessory set (6.1602.160)
This accessory set contains the bottle cap, an M6 tubing nipple, an M8
tubing nipple, two O-rings and an M6 and M8 threaded stopper.
■ The tubing adapter for aspiration filter accessory set (6.2744.210)
This accessory set contains a filter holder, a clamping screw and tubing
weighting.
■ An aspiration filter (6.2821.090)
■ The adsorber tube (6.1609.000)
■ The SGJ clip (6.2023.020)
Connecting the eluent aspiration tubing
1
Installing the eluent bottle cap (6.1602.160)
■ Start by pushing the M8 tubing nipple onto the loose end of the
eluent aspiration tubing, followed by the O-ring.
■ Push the loose end of the eluent aspiration tubing through the
M8 opening of the bottle cap and screw it on for the time being.
Figure 7 Installing the eluent bottle cap