4 Installation
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
■■■■■■■■
27
4.8 Installing an inline filter
Inline filters protect the separation column reliably from potential contam-
ination from the eluent. The small filter pads with 2 µm pore size can be
replaced quickly and easily. They remove particles from the solutions.
An inline filter (6.2821.120) is installed between the purge valve and the
pulsation absorber as protection against particles.
1
2
6.2744.070
6.2744.070
6.2821.120
Figure 10 Inline filter
1
Inlet capillary
Connected to the purge valve.
2
Outlet capillary
Connected to the pulsation absorber.
The inline filter is completely connected. No installation work is required.
4.9 Installing the pulsation absorber
The pulsation absorber is installed between the high-pressure pump and
the injection valve. It protects the separation column from damage caused
by pressure fluctuations, e.g. when the injection valve is switched, and
reduces interfering pulsations during highly sensitive measurements.

Table of Contents

Other manuals for Metrohm 930 Compact IC Flex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 930 Compact IC Flex and is the answer not in the manual?

Metrohm 930 Compact IC Flex Specifications

General IconGeneral
BrandMetrohm
Model930 Compact IC Flex
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals