4.3 Removing transport locking screws
■■■■■■■■■■■■■■■■■■■■■■
18
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
4.3 Removing transport locking screws
To avoid damage to the drives for the high-pressure pump and the vac-
uum pump during transport, the pumps are secured with transport locking
screws. These are located at the rear of the instrument and labeled with
Transport security screws.
Remove these transport locking screws before the initial start-up.
Accessories For this step you need:
■ 4 mm hex key (6.2621.030)
Removing the transport locking screws
Figure 6
Removing the transport locking screws
1
Transport locking screws
For the vacuum pump. Only present if the
instrument has a degasser or a CO
2
Suppres-
sor (MCS).
2
Transport locking screws
For the high-pressure pump.
1
Remove all of the transport locking screws with the hex key.