6 Operation and maintenance
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
■■■■■■■■
81
Inserting a new filter
1
Inserting a new filter
Use tweezers to carefully place a new filter into the filter screw
(23-2) so that it is flat and press it firmly into place with the rear
of the tweezers.
2
Installing the filter screw
Screw the filter screw (23-2) back into the filter housing (23-1)
and tighten by hand. Then use two adjustable wrenches
(6.2621.000) to tighten it slightly.
3
Remounting the inline filter
Screw pressure screws back onto the inline filter.
Take care to ensure that the flow direction matches that inscribed
on the inline filter.
4
Rinsing the inline filter
Dismantle the guard column (if present) and the separation col-
umn and replace with a coupling (6.2744.040).
Rinse the instrument with eluent.
Reinsert the columns after 10 minutes.