Troubleshooting
After releasing the option doesn‘t work
Check if the option is released. Use command ge
toptions.
Before you try again to release the option, please make sure
that your release code is correct. A wrong code will disable all
options.
In most cases you got the code via
E-mail. It is a reasonable
way to copy the number from the E-mail into the clipboard and
insert it directly in the Venus command line.
- Don‘t forget to send command save after setc
ode.
- Don‘t forget to send command reset after s
ave.
During releasing with WinPos you got the message
"unknown command setcode"
You are using an older WinPos Version. Ask you dealer for
WinPos Version 1.1 with "venuscom.dat"
date code: 14.07.2003 or newer.