Appendix A AccuPyc II 1340
A-8 Oct 2013
MAN_ERR:
Pressure overrange
No A.C. Found
No collected data to report,
or all cycles excluded
Printer port
not responding
Queuing print job
[ESCAPE] to cancel
Cause: The fill valve was left open until the maximum system pressure was
exceeded.
Action: Close the fill valve and open the vent and expansion valves. Allow the
pressure to stabilize.
Cause: The control module cannot detect an analysis module.
Action: Contact your Micromeritics service representative.
Cause: You requested a report in which there is either no data available or data
have been excluded via review mode.
Action: Initiate an automatic operation or return to review mode and include at
least one cycle.
Cause: Your printer cable is not connected or the printer is not turned on.
Action: Check to make sure the printer is properly connected to the
pycnometer, is turned on, and is on line.
Cause: Status message displayed when a report is being printed.
Action: None; this is a status message only.