AccuPyc II 1340 Appendix A
Oct 2013 A-9
Reset to factory
defaults
Sending line (line number)
[ESCAPE] to cancel
Transmission port
not responding
Transmission port
waiting for Xon
Cause:
The
. (period) key was pressed at startup and the instrument was reset.
Data files and HTML reports have been erased. Setup options have
been returned to default values.
Action: Review setup options.
Cause: Status message displayed when data are being transmitted.
Action: None; this is a status message only.
Cause: The receiving device took longer than five seconds to acknowledge
receipt of data from the pycnometer.
Action: Make sure the receiving device is properly connected to the
pycnometer RS-232 port and is turned on. Verify that the serial I/O
parameters controlling the receiving device correspond with the data
transmission parameters in the set up mode.
Cause: The receiving device stopped transmission by sending an Xoff, and
hasn’t resumed the transmission by sending an Xon.
Action: None; when the receiving device is ready for more data, it should send
the pycnometer an Xon.