System Commands AccuPyc II 1340
3-14 Aug 2013
Report Options
The Report options command enables you to specify the report mode (density or volume for analysis)
and the report destination.
This prompt displays when you selected Yes for run precision
and if you have requested more than five cycles.
Enter the run precision volume tolerance which is expressed
as a percent of nominal cell volume (sample capacity), then
press
ENTER.
Range: 0.01% to 50%
Press
CHOICE until Report Options is displayed, then press
ENTER.
Press
CHOICE to select Density or Volume, the mode in
which the analysis report will be displayed on the
pycnometer.
If you select Density, a Sample Mass? prompt displays
during the Analysis function so that you can enter a value for
the sample mass. For accurate data, the sample mass must be
entered when choosing Density.
Press
ENTER.
A sample ID (identification) is a unique identifier of the
sample. Using sample IDs can help you keep track of data
from multiple analyses. For example, you can use the
sample ID as a date and time stamp. A Sample ID prompt
displays during the analysis operation, allowing you to enter
the identification when you select Yes to this prompt.
Choices: Yes, No
Press
CHOICE to select Ye s or No, then press ENTER.
Unit[n] SN1234 10 cm3
Analysis Parameters
Percent Full-Scale:
0.05%
Unit[n] SN1234 10 cm3
Set-up Type?
Report Options
Unit[n] SN1234 10 cm3
Report Options
Anls Display Mode?
Density
Unit[n] SN1234 10 cm3
Report Options
Request Sample ID?
Yes