AccuPyc II 1340 System Commands
Aug 2013 3-31
Calibration
Press Alt + 5 to review or edit calibration results; the following prompts display if they were selected
for the calibration. You may edit the information in the prompts if desired
Displays the Density or Volume (depending on the selection
you made for Analysis display mode).
[n] shown in the prompts to the left represent the cycle num-
ber.
Press
CHOICE to exclude the displayed density (or volume)
from the calculated average. This will place an asterisk next
to it indicating that it has been excluded. Press
CHOICE
again to remove the asterisk and have it included.
Each time you press
CHOICE to exclude (or include) the
value, a new deviation is calculated and displayed.
Press
ENTER to view the values for the next cycle.
Press
SAVE when you wish to return to the Reload prompt.
When you press
SAVE, all collected data are recalculated
automatically, and all data reduction messages are added
back into the queue.
Displays the insert you selected for the calibration or None
if an insert was not used. If the selection is incorrect, press
CHOICE until the correct one is shown, then press ENTER.
Choices: None, 0.1 cm
3
(for 1 cm
3
pycnometer)
None, 3.5 cm
3
, 1.0 cm
3
(for 10 cm
3
pycnometer)
None, 35 cm
3
, 10 cm
3
(for 100 cm
3
pycnometer)
None, 650 cm
3
, 1300 cm
3
(for 2000 cm
3
pycnometer)
Displays the volume of the calibration standard that was
used for the calibration.
Press
ENTER.
Unit[n] SN1234 10 cm3
X - X - X
Dn [n] = (density)
Dv [n] = (deviation)
or
Unit[n] SN1234 10 cm3
X - X - X
V [n] = (volume)
Dv [n] = (deviation
Unit[n] SN1234 10 cm3
X - X - X
Chamber insert?
(insert)
Unit[n] SN1234 10 cm3
X - X - X
Volume of Cal Std:
(volume)