System Commands AccuPyc II 1340
3-32 Aug 2013
Enables you to choose whether you wish to view calibration
data for the sample cell chamber or the expansion chamber.
Choices: Cell volume, Expansion volume
Press
CHOICE until the desired chamber is displayed, then
press
ENTER.
Displays the time and date the calibration began.
Press
ENTER.
Displays the time and date the calibration completed.
Press
ENTER.
Displays the cell or expansion volume (depending on the
selection you made at the Which Chamber prompt), and
the deviation from the mean.
[n] shown in the prompts to the left represent the run num-
ber.
Press
CHOICE if you wish to exclude the displayed density
(or volume) from the calculated average. This will place an
asterisk next to it indicating that it will not be included.
Press
CHOICE again to remove the asterisk and have the
value again included in the average.
Each time you press
CHOICE to exclude (or include) the
value, a new deviation is calculated and displayed.
Press
ENTER to view the values for the next cycle.
Press
SAVE when you wish to return to the Reload prompt.
When you press
SAVE, all collected data are recalculated
automatically, and all data reduction messages are added
back into the queue.
Unit[n] SN1234 10 cm3
X - X - X
Which Chamber?
(chamber type)
Unit[n] SN1234 10 cm3
X - X - X
Start:
(HH:MM:SS) (DD/MM/YY)
Unit[n] SN1234 10 cm3
X - X - X
End:
(HH:MM:SS) (DD/MM/YY)
Unit[n] SN1234 10 cm3
X - X - X
Cell [n] = (volume)
Dev [n] = (deviation)
or
Unit[n] SN1234 10 cm3
X - X - X
Exp [n] = (volume)
Dev [n] = (deviation