Maintenance
114
Routine checks
Before the start of each working day the user must carry out a
number of routine checks. A check list is supplied with the machine
for this purpose.
The following need to be inspected:
All filters in the wash cabinet
The spray arms in the machine and in any mobile units or baskets
The wash cabinet and the door seal
The dispensing systems and
Mobile units, baskets, modules and inserts.
Cleaning the filters in the wash cabinet
The filters in the floor of the wash cabinet prevent coarse soiling from
coming into contact with the circulation system. Filters can become
blocked by soiling, so they need to be checked every day and
cleaned as necessary.
This machine must not be used without all the filters in place.
In the controls, it is possible to set a cleaning interval for the filters in
the wash cabinet, see "Settings/Filter maintenance".
The cleaning interval is not a substitute for the daily routine check of
the filters in the wash cabinet!
Danger of injury from glass shards, needles, etc. retained in the
filters.
Turn the microfine filter in the direction of the arrow and remove it
together with the coarse filter.