Areas of application
40
Only reprocess small items and micro components in special
inserts, mesh trays with lids or mesh inserts.
Heat-sensitive items must only be reprocessed using a chemo-
thermal programme.
For validation purposes it is essential to follow the loading
instructions given on the template.
Observe the further information given in the following sections as
necessary depending on area of application.
Preparing the
load
Empty the load before sorting
Ensure that no acid or solvent residues, especially hydrochloric
acid or chlorides, get inside the wash cabinet.
Dismantle the items for washing where possible according to the
manufacturer's instructions and open any valves or taps.
Follow the instructions of the manufacturer regarding pre-cleaning
and pre-treatment as necessary.
Thoroughly rinse items which have been pre-treated with chemicals
(see "Wet loading")
Dry loading
Contaminated medical products should be placed directly into
baskets and inserts in the washer-disinfector after use without pre-
treatment
Dry loading is preferable for contaminated medical items.
Wet loading
Chemically pre-treated items must be rinsed thoroughly by hand or
using the Rinsing programme before reprocessing in the washer-
disinfector to avoid a significant build-up of foam.