INSTALLATION
- 11 -
4-4 INSTALLATION OF THE SYNERGYPAK
Do not touch the UV Lamp when replacing the
SynergyPak.
HAZARD
!
INSTALLATION
Make sure the front cover is opened.
STANDBY should be viewed on the Display
(A).
A
B
Remove the SynergyPak from its shipping box.
Remove the protective caps on the ports of the
SynergyPak and from the system.
Locate the O-rings on the ports (B). Wet them
with water. It is preferable to wet them with
ultrapure water.