INSTALLATION
4-6 PURGING AIR FROM THE SYSTEM
A
At this time you should have installed the
SynergyPak, Barbed Fitting and the Tygon
Tubing. Air trapped in the SynergyPak should
now be purged from the system.
Press the Dispensing Button once to put the
system into DISPENSING mode (A).
Dispense about 3-4 Litres of water from the
system.
4-7 HYDRATING THE SYSTEM
At this time, the Final Filter is not installed.
Leave the system overnight or for several hours
in PRE OPERATE mode (see
Section 5-3
Operating Modes
). The system will regularly
recirculate water and rinse off purification
media inside the SynergyPak. Do not leave the
system in STANDBY mode.
(The next morning) Press the Dispensing Button
once to put the system into DISPENSING
mode.
Allow about 1 Litre of water to be dispensed
from the system.
Press the Dispensing Button once again. The
system will finish dispensing water.
Remove the Tygon Tubing and the Barbed
Fitting from the POU Dispenser.
The Final Filter needs to be installed. See
Section
4-8 Installation and Rinsing of the Final Filter.
- 14 -