USING THE SYNERGY
- 28 -
5-6 HOW TO UNDERSTAND SYNERGY MESSAGES
PACK ALARM
The system will prompt you to change the SynergyPak using a red
Pack Alarm icon. The Display will show the red Pack Alarm
blinking. The SynergyPak is changed due to either the amount of
time it has been used or from the amount of water that has passed
through it.
When the Resistivity display is blinking, the red Pack Alarm will
also be blinking. This indicates that the SynergyPak should be
replaced. This message is shown when the Product resistivity is
less than the Resistivity Setpoint. The Resistivity Setpoint can be
seen in menu option C02
When the red Pack Alarm is displayed as a steady icon, the
SynergyPak is not installed correctly or not installed at all. If the
SynergyPak has been reinstalled and the Alarm is still displayed,
then contact Millipore.
UV
LAMP ALARM
The system will prompt you to change the UV Lamp using a
red UV Lamp Alarm icon. The Display will show the red UV
Lamp Alarm blinking. The message is shown when the UV
Timer displays 0 days. The UV Timer can be viewed in menu
option C03. The replacement of the UV Lamp involves the
installation of a new UV Lamp and a manual reset of the UV
Timer.
When the red UV Lamp Alarm is displayed as a steady icon,
the UV Lamp is not installed correctly or not installed at all. If
the UV Lamp has been reinstalled and the Alarm is still
displayed, then contact Millipore.