MAINTENANCE
6-6 HOW TO VIEW OR RESET THE UV LAMP TIMER (C03)
The UV Lamp Timer should be reset only after the UV Lamp has been replaced (see
Section 6-5 How to
Replace the UV Lamp
). The UV Lamp Timer displays the time left until the UV Lamp needs to be replaced.
The Display will show the red UV Lamp Alarm icon blinking when the Timer reaches 0 days. This
message is displayed until the UV Lamp is replaced and the UV Lamp Timer is reset.
H
OW TO VIEW THE REMAINING DAYS ON THE UV LAMP TIMER
1. Press the Main and
“+“ Buttons together to
enter the menu. The
Display will show
C01.
2. Press the Main Button
2 times. The Display
will show C03 and the
days left on the UV
Timer.
3. To exit menu, press
and hold the Main
Button for 2 seconds.
To display the next
menu option, press the
Main Button once.
- 38 -