INSTALLATION
4-9 HOW TO CALIBRATE THE FLOWRATE FROM THE SYNERGY (F02)
Before calibrating the Product flowrate from the Synergy, you will need a 1 Litre graduated
cylinder to measure the total volume of water that will be dispensed. The Final Filter should be
installed.
1. Press the Main and
“-“ Buttons together
to enter the menu.
The Display will
show F01.
2. Press the Main Button
once. The Display
will show F02 and
the 60 second timer.
The graduated
cylinder will be
blinking.
3. Press the Dispensing
Button once. The
system will dispense
water for 60
seconds. The Display
will show the timer
counting down.
CONTINUED ON NEXT PAGE