INSTALLATION
4-8 INSTALLATION AND RINSING OF THE FINAL FILTER
Obtain a Final Filter. It can be the Millipak Express 20 or the BioPak Ultrafiltration Cartridge.
Remove the Tygon Tubing and the Barbed Fitting from the POU Dispenser.
Screw the Final Filter onto the end of the POU Dispenser. The Final Filter should be turned until it is
hand tight (A). Do not over tighten the Final Filter.
ATTENTION
Do not use white tape on the threads of the
Final Filter. The POU Dispenser has an O-ring
inside which provides a watertight seal.
Press the Dispensing Button once. This will bring the system into DISPENSING mode.
(If your Final Filter is a Millipak)
Locate the air vent valve on the top side of the Millipak. Open this
slowly but do not remove it from the Millipak (B). Allow any air on the clear side of the Millipak to
be vented out.
Air will not pass through the membrane
filter in the Millipak. If there is trapped air
in the Millipak, then a reduction in Product
Water flowrate can result.
ATTENTION
Dispense about 1 Litre of water. Make sure all air is purged out.
Press the Dispensing Button once again. The system will finish dispensing water.
Leave the system in PRE OPERATE mode.
It is highly recommended not to put Tygon
Tubing or any other type of tubing on the
end of the Final Filter. This can compromise
the Product Water quality (bacteria can grow
in the Tygon Tubing).
ATTENTION
The Water System is now ready for use.