Using the QC Programs
8-20
z When switching from the Predilute mode to the Whole Blood mode, the
analyzer will automatically flush the fluidic system.
z Be sure to keep the sample probe tip away from the tube bottom, otherwise
the aspiration volume may be inaccurate.
z When the aspiration is done, remove the sample tube only when the sample
probe is out of the tube.
Running the controls
1 Be sure the System Status area displays “Ready“ and Count Mode area displays
Whole“;
2 Present a vial of control to the sample probe so that the tip is well into the vial, and press
the aspirate key. The System Status area will display “Running” and the analyzer will
start aspirating control;
3 When you hear the beep and the sample probe is out of the vial, remove the control vial.
The sample probe will retract into the analyzer and the analysis progress will be
displayed on the screen;
4 When the analysis is finished, the sample probe is replaced, the analysis result is
displayed on the screen, and a message box pops up to confirm the validity of the
analysis result, as Figure8-24 shows;
Figure8-24 A message to confirm the validity
5 CLICK Enter” to save the result and the “NO./Total” in the upper left corner of the
screen will automatically increase by 1; CLICK Cancel” to abort the result;
6 Follow the above steps to have another QC run. When you have obtained two valid QC
results, the analyzer will calculate the average and take it as an
X
analysis result. The
average will be flagged “H” or “L” if it falls outside the expected range.