Using the QC Programs
8-25
The “” points represents non-numeric parameter values (*).
If you see any points fallen outside the control range, do the following steps until the problem
is solved. If all the steps have failed, contact Mindray customer service departmentor your
local distributor for assistance.
1. Check the upper left corner of the screen for error messages. Refer to Chapter 11
Troubleshooting Your Analyzer for solutions to any displayed error messages;
2. Check the
X
settings for inappropriate entries;
3. Do the background check. In case of an abnormal background result, refer to
Chapter 11 Troubleshooting Your Analyzer for solutions;
4. Re-run the control;
5. Run another vial of control;
6. Check if the analyzer needs to be calibrated.
Browsing results of
X
analyses
Press [] or []to review the preceding or following screen; press[] or [] to review the
preceding or following result. The parameter value of the current point (the one the cursor is
located at) is displayed below the parameter box. The location of the current point is
displayed in the “No.” field. The analysis time is displayed in the “Timefield.
Printing
X
graphs
Press [PRINT] to print out the displayed
X
graphs.
Exiting the “
X
Graph” screen
Press [MENU] to exit to the system menu, or press [MAIN] to exit to the “Count” screen.
X
Table” mode
Entering the “
X
Tablemode
Press [MENU] to enter the system menu.