Using the Calibration Programs
9-11
Figure 9-8 A message box to confirm the new calibration factors
CLICK ”Enter” to save the new calibration factors to the “Manual Calibration” screen and
enter the “Count” screen.
At the “Count” screen, run the calibrator or a normal control material at least 5 consecutive
times and calculate the mean of the results. The means should be within the expected ranges
supplied by the manufacturer. If not, contact Mindray customer service department or your
local distributor for assistance.
Printing new calibration factors
Press [PRINT] to print out the new calibration factors.
Exiting the “Auto Calibration” screen
Press [MENU] to exit to the system menu or [MAIN] to exit to the “Count” screen. A message
box will pop up to confirm the new calibration factors, as Figure 9-8 shows. CLICK “Enter” to
save the new factors to the “Manual Calibration” screen and exit to the system menu or the
“Count” screen; CLICK “Cancel” to abort the new factors and exit to the system menu or the
“Count” screen.