Using the Calibration Programs
9-12
9.3.3 “Fresh Blood” Program
Press [MENU] to enter the system menu.
Figure 9-9 System menu
SELECT Calibration Fresh Blood” (Figure 9-9) to enter the “Fresh Blood” screen
(Figure9-10).
Figure9-10 Fresh Blood screen
Complete the fresh blood calibration as instructed below:
Selecting count mode
Press [MENU] and SELECT Mode” to enter the “Mode” screen. SELECT Whole Bloodor
Predilutefrom the Sample Mode pull-down list.