EasyManua.ls Logo

Mindray BC-3000 Plus - Page 230

Mindray BC-3000 Plus
325 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using the Calibration Programs
9-14
Figure9-12 A message box to confirm the validity
CLICK Enter” to save the results to the “Auto – fresh blood” screen; CLICKCancel” to
abort the result.
z If you press [MENU] to enter the system menu before the average is
obtained at the “Calculate” screen, the next time you enter the “Auto-fresh
blood” screen a message box will pop up to ask you whether to clear the
data of the last calibration; If you press [MAIN] to exit to the “Count”screen
before the average is obtained at the “Calculate” screen, the next time you
enter the “Auto-fresh blood” screen a message box will pop up to ask you
whether to clear the data of the last calibration
Repeat the above steps to run the sample 3 – 5 times (5 is recommended) and the analyzer
will automatically calculate the CV and calibration factor, as Figure9-13 shows. Be sure the
CVs meet the requirements of Table 9-1.
Figure9-13 “Fresh Blood” screen

Table of Contents

Related product manuals