EasyManua.ls Logo

Mindray BC-3000 Plus - Page 232

Mindray BC-3000 Plus
325 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using the Calibration Programs
9-16
Figure 9-15 A message box to confirm the new calibration factors
CLICK Enter” to save the new calibration factors to the “Manual Calibration” screen and
enter the “Count” screen.
Test the new calibration factors either of the following ways.
Method one:
Prepare 3-5 normal fresh blood samples and run each one of them on a reference analyzer at
least 3 consecutive times. Calculate the mean (MEAN 1) and SD (SD 1) of every sample.
Run the same samples on your analyzer for the same number of times and calculate the
mean (MEAN 2). The MEAN 2 should be within MEAN 1 ± 2SD. If any of the sample fails to
reach the criterion, call Mindray customers service department or your local distributor.
Method two:
At the “Count” screen, run the calibrator at least 5 consecutive times and calculate the
means of the results. The means should be within the expected ranges supplied by the
manufacturer. If not, contact Mindray customer service department or your local distributor.
Printing new calibration factors
Press [PRINT] to print out the new calibration factors.
Exiting the “Fresh Blood” screen
Press [MENU] to exit to the system menu. A message box will pop up to confirm the new
calibration factors, as Figure 9-15 shows. CLICKEnter” to save the new factors to the
Manual Calibrationscreen and exit to the system menu; CLICKCancel” to abort the new
factors and exit to the system menu.
9.3.4 Manual Calibration Program
If needed, you may run the calibration material at the “Count” screen and calculate the
calibration factors manually.
Press [MENU] to enter the system menu. SELECT Count” (Figure7-1) to enter the “Count
screen (Figure9-17).

Table of Contents

Related product manuals