Troubleshooting Your Analyzer
11-13
5. Enter the “Setup Count Time
screen and record the RBC count
time. Then enter the “Service
System Test” screen and test the
actual RBC count time as
instructed by Chapter 10.6.;
6. If the difference between the
reference RBC count time and the
actual RBC count time is less than
2 seconds, the error has been
removed.
7. If the difference is still greater than
2 seconds but consistent, enter the
Setup Count Time” and reset
the RBC count time. Then enter the
Service System Test” screen
and test the actual RBC count
time as instructed by Chapter 10.6
to confirm the difference is less
than 2 seconds.
8. If the problem remains, contact
Mindray customer service
department or your local distributor.
RBC bubbles
1. Diluent or rinse running
out;
2. Loose tube connections;
3. Inappropriate RBC
count time setting.
1. Check if the diluent or rinse has run
out. If so, change a new container
of diluent or rinse as instructed in
Chapter 4.4.2;
2. Check the connection of the diluent
and rinse pickup tube. If necessary,
reconnect and tighten them as
instructed by Chapter 4.4.2;
3. If the problem remains, adjust the
RBC count time as instructed by