6 - 7
Battery Error Low Battery fault, such as
battery over heat,
charging failure, etc.
Contact your service personnel.
CMS/eGW
Disconnected
Low The pump is
disconnected from the
CMS, the wireless
network connection
symbol disconnects.
Reconnect the pump with
the central station, the
wireless network connection
symbol restores.
If the alarm persists, contact
your service personnel.
Standby Time
Expired
Low The preset standby time
is reached.
Press to reset the alarm.
Dock
Connection
Interrupted
Low The pump is
disconnected from the
Dock.
Reinsert the pump to the
Dock.
Insert the pump to another
pump bay.
If the alarm persists, contact
your service personnel.
System Time
Error
Low The real time clock (RTC)
reset or RTC fault.
Re-set the system time. See
3.4.4 Setting the Date and
Time.
If the alarm persists, contact
your service personnel.
Relay Invalid
Soon
Low In the relay state, the
upstream pumps have
almost completed the
infusions yet the
downstream pumps are
not ready for infusion.
Check that the downstream
pumps are properly configured
for infusion.
Time Near
End
Low The remaining infusion
time reaches the
configured time near
end.
Complete the infusion or
prepare for a new therapy.
Reminder Low No operation is detected
after the preset
Reminder Time is
reached.
Turn off the pump or enter the
standby.
Low Battery Low Low battery. Connect the pump to the
external power source.
Alarm Priority Causes Solutions