13 - 2 BeneVision N Series Patient Monitor Operator’s Manual
13.3 Resp Display
NOTE
If ESU-proof ECG cables are used, the Resp waveform area will display the message Check Leads.
Replace the ECG cable if necessary.
13.4 Preparing for Resp Monitoring
13.4.1 Preparing the Patient
Follow this procedure to prepare the patient:
1. Shave hair from skin at chosen sites.
2. Gently rub skin surface at sites to remove dead skin cells.
3. Thoroughly cleanse the site with a mild soap and water solution.
4. Dry the skin completely before applying the electrodes.
CAUTION
Proper skin preparation is necessary for good signal quality at the electrode site, as the skin is a poor
conductor of electricity.
(1) Resp waveform gain (2) Resp lead label (3) Alarm limits
(4) Respiration rate (RR) (5) RR source
(1)
(5)
(4)
(3)
(2)