EasyManuals Logo

Mindray BeneVision N Series User Manual

Mindray BeneVision N Series
528 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #333 background imageLoading...
Page #333 background image
BeneVision N Series Patient Monitor Operator’s Manual 34 - 5
34.1.5.5 Setting Thresholds for Triple Low Parameters
To set low limits for triple low parameters, follow this procedure:
1. Select
.
2. Select the Maintenance tab → Triple Low tab.
3. Set low limits for BIS, MAC, and MAP.
34.2 Early Warning Score (EWS)
The Early Warning Scores (EWS) can help you recognize the early sign of deterioration in patients based on vital
signs and clinical observations. Depending on the score calculated, appropriate recommendations are
displayed.
The monitor supports the following scores:
â–  MEWS (Modified Early Warning Score)
â–  NEWS (National Early Warning Score)
â–  NEWS2 (National Early Warning Score 2)
â–  Custom Score
There are two types of scoring tools:
â–  Total score: A subscore is given for each parameter based on the measured or entered value. When all the
required parameters are entered or measured, the subscores are added together to calculate the total early
warning score. Each subscore has a color coding to indicate associated level of risk, When the total score is
outside of the thresholds, actions are recommended. MEWS, NEWS and NEWS2 can give total scores.
â–  IPS (individual parameter score): A color-coded score is given for each parameter based on the measured or
entered value. Each parameter has upper and lower thresholds. When an individual parameter measured
or entered is outside of the thresholds, actions are recommended.
Custom Score is based on user-defined parameters. It can be a total score or an IPS, depending on the
configuation.
MEWS, NEWS and NEWS2 are intended for adult patients only. The patient category applied to the Custom Score
is defined by Mindray Clinical Score Configuration Tool. For more information, see Mindray Clinical Scoring Config
Tool Instruction for Use (P/N: 046-007126-00).
WARNING
• The EWS scores and recommended actions are for reference only and cannot be directly used for
diagnostic interpretation.
• EWS cannot be used as an prognosis index. It is not a clinical judgement tool. Clinicians must use
their clinical judgement in conjunction with the EWS tool at all times.
• MEWS and NEWS are not applicable to pregnant woman, COPD (Chronic Obstructive Pulmonary
Disease) patients and patients under 16 years old. NEWS2 is not applicable to pregnant woman and
patients under 16 years old.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals