14 - 8 BeneVision N Series Patient Monitor Operator’s Manual
14.6.9 Changing the Sweep Speed of the Pleth Wave
To set the sweep speed of Pleth waveforms, follow this procedure:
1. Select the SpO
2
numeric area or waveform area to enter the SpO2 menu.
2. Select the SpO2 Setup tab.
3. Set Speed.
14.7 Changing the PR Settings
14.7.1 Changing the PR Alarm Settings
To change the PR alarm settings, follow this procedure:
1. Select the SpO
2
numeric area or waveform area to enter the SpO2 menu.
2. Select the PR Alarm tab.
3. Enter the password if required. For more information, refer to 39.13The Authorization Setup Settings.
4. Set the alarm properties as desired.
14.7.2 Changing the QRS Volume
If the Alarm Source is set to PR, the QRS tone is derived from PR measurements. To set the QRS volume, follow
this procedure:
1. Select the SpO
2
numeric area or waveform area to enter the SpO2 menu.
2. Select the PR tab.
3. Select the Setup tab.
4. Set QRS Volume.
If the SpO
2
value is effective, the monitor also adjusts the QRS tone (pitch tone) according to the SpO
2
value.
14.7.3 Setting the PR Source
Current pulse source is displayed in the PR numeric area. The PR from current pulse source has the following
characteristics:
â–  PR is monitored as system pulse and generates alarms when you select PR as the active alarm source.
■ PR is stored in the monitor’s database and reviewed in the graphic/tabular trends; in trend graphs, as the PR
curve is in the same color with that of the PR source, it is unlikely to distinguish the PR source.
â–  PR is sent via the network to the CMS, if available.
To set which pulse rate as PR source, follow this procedure:
1. Select the SpO
2
numeric area or waveform area to enter the SpO2 menu.
2. Select the PR Setup tab.
3. Set PR Source.
The PR Source menu displays the currently available PR sources from top to bottom by priority. When you select
Auto, the system will automatically select the first option as the PR source. If the current PR source is unavailable,
the system will automatically switch PR Source to Auto. When you select IBP, the system will automatically
select the first pressure label as the PR source.
14.7.4 Showing/Hiding PR
You can set whether to display the PR value in the SpO
2
parameter area. To do so, follow this procedure:
1. Select the SpO
2
numeric area or waveform area to enter the SpO2 menu.
2. Select the PR tab.
3. Select the Setup tab.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals