BeneVision N Series Patient Monitor Operator’s Manual 5 - 3
4. Select a patient from the patient list, and then select Import. Corresponding patient information in the
monitor will be updated.
NOTE
• You can load patient information from the ADT server only when ADT Query is enabled. For more
information, see 7.5Viewing Other Patients.
• Loading patient information from the ADT server updates only patient information in the monitor.
The patient’s monitoring data is not changed and the patient is not discharged.
5.5 Transferring Patient
You can transfer a patient via the BeneVision N1(hereafter referred to N1), BeneView T1 (hereafter referred to T1),
or the MPM module (hereafter referred to MPM), to another monitor without re-entering the patient
demographic information or changing the parameter settings. Transferring of patient data enables you to
understand the patient’s history condition.
WARNING
• Do not discharge a patient before the patient is successfully transferred.
• Do not remove the N1/ T1/MPM from the monitor before parameter settings are synchronized
between N1/ T1/MPM and the monitor (this takes maximum 30 seconds). Otherwise, patient
information and measurement data saved in the N1/ T1/MPM may not be consistent with those in
the monitor.
• Removing the N1/ T1/MPM when transferring historical patient data to the monitor will cause
historical data saved in the monitor incomplete.
• After a patient is successfully transferred, check if the patient settings (especially patient category,
paced status, alarm limits settings, and etc) on the monitor are appropriate for this patient.
NOTE
• The system automatically switches on the HR alarm and lethal arrhythmia alarm after transferring
the patient data.
5.5.1 Data Storage Introduction
Understanding the data respectively stored in the this monitor, N1, T1, or MPM helps you understand the effects
incurred by transferring patients with an N1, T1, or MPM.
Type of storage Can be stored
in the monitor?
Can be
transferred via
MPM?
Can be
transferred via
T1?
Can be
transferred via
N1?
Data Patient demographics Yes Yes Yes Yes
Trend data Yes yes Yes Yes
Calculation data Yes No No Yes
Event data Yes No Yes (only data
monitored by
T1)
Yes (data
monitored by
both N1 and the
monitor)
Full disclosure Yes No No Yes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals