BeneVision N Series Patient Monitor Operator’s Manual A - 3
A.3 Environmental Specifications
WARNING
• The monitor may not meet the performance specifications if stored or used outside the specified
temperature and humidity ranges. If the performance of the equipment is degraded due to aging or
environmental conditions, contact your service personnel.
• When the monitor and related products have differing environmental specifications, the effective
range for the combined products is that range which is common to the specifications for all
products.
NOTE
• The environmental specification of unspecified parameter modules are the same as those of the
main unit.
Components Item Operating Condition Storage Condition
Main Unit Temperature (ºC) 0 to 40
(0 to 35 for N17 configured
with iView module)
-20 to 60
Relative humidity
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 427.5 to 805.5 120 to 805.5
Microstream CO
2
module
Temperature (ºC) 0 to 40 -20 to 60
Relative humidity
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 430 to 790 430 to 790
Sidestream CO
2
module
Temperature (ºC) 5 to 40 -20 to 60
Relative humidity
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 430 to 790 430 to 790
Mainstream CO
2
module
Temperature (ºC) 0 to 40 -20 to 60
Relative humidity
(noncondensing) (%)
10 to 90 10 to 90
Barometric (mmHg) 427.5 to 805.5 400 to 805.5
AG module Temperature (ºC) 10 to 40 -20 to 60
Relative humidity
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 525 to 805.5 525 to 805.5
RM module Temperature (ºC) 5 to 40 -20 to 60
Relative humidity
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 427.5 to 805.5 120 to 805.5
ICG module Temperature (ºC) 10 to 40 0 to 50
Relative humidity
(noncondensing) (%)
15 to 95 15 to 95
Barometric (mmHg) 427.5 to 805.5 120 to 805.5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals