6 - 2 BeneVision N Series Patient Monitor Operator’s Manual
■ A patient is discharged.
■ Patient data is cleared.
■ Patient category is changed.
To set the default configuration, follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select Manage → input the required
password → select .
2. Select Select Default Config.
3. Select Load the Latest Config or Load Specified Config.
◆ When you select Load Specified Config, the restored configuration is subject to the patient category
(adult, pediatric or neonate). This configuration can be either factory configuration or a saved user
configuration. As an example, select Default Adult Config and then select Factory Default or user
configuration(s).
◆ When you select Load the Latest Config, the latest configuration is loaded when the monitor is
started or a patient is admitted.
6.5
Saving Current Settings
C
urrent settings can be saved as a user configuration. Up to 25 user configurations can be saved.
To save current settings, follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select Manage → input the required
password → select .
2. Select Save Current Settings.
3. Input the configuration name.
4. Select Ok to save current settings as a user configuration.
6.6 Deleting a Configuration
To delete a configuration, follow this procedure:
1. Select the Main Menu quick key → from the Configuration column select Manage → input the required
password → select .
2. Select Delete Configuration.
3. Select the configuration you want to delete:
◆ In the Delete Configuration menu, selecting Local tab shows the existing user configurations on the
monitor.
◆ In the Delete Configuration menu, selecting USB Drive tab shows the existing user configurations
on the USB drive.
4. Select Delete.
5. Select Ok.
6.7 Transferring a Configuration
When installing several monitors with identical user configurations, it is not necessary to set each unit
separately. Use a USB drive to transfer the configuration from monitor to monitor.
6.7.1 Exporting a Configuration
To export the current monitor’s configuration, follow this procedure:
1. For N17/N15/N12/N12C, connect the USB drive to the monitor’s USB connector. For N22/N19. connect the
USB drive to the monitor’s MSB connector.
2. Select the Main Menu quick key → from the Configuration column select Manage → input the required
password → select .