A - 22 BeneVision N Series Patient Monitor Operator’s Manual
A.15.11 CCO Specifications
A.15.11.1 CCO Specifications from Vigilance II, Vigileo, and EV1000 monitor
ScvO
2
0 to 99% 50% to 80%: ±3%
Other ranges: Not specified.
Alarm Limit Range Step
ScvO
2
High ScvO
2
<60%: (Low limit + 5%) to 60%
ScvO
2
≥60%: (Low limit + 1%) to 99%
ScvO
2
<60%: 5%
ScvO
2
≥60%: 1%
ScvO
2
Low ScvO
2
<60%: 0 to (High limit - 5%)
ScvO
2
≥60%: 60% to (High limit - 1%)
Operating mode Interfaces with Edwards Vigilance II, Vigileo, or EV1000 monitor
Measured parameter Consistent with CCO-related parameters outputted by Vigilance II, Vigileo, and
EV1000 monitor
Parameter alarm CCO, CCI
Signal Outputs
Standard Meets the requirements of IEC 60601-1 for short-circuit protection and
leakage current
Output impedance ≤1000
Isolation voltage 1500 VAC
ECG Analog Output
Bandwidth (-3dB; reference frequency:
10Hz)
ST mode: 0.05~40Hz
Diagnostic mode: 0.05~150Hz
Monitor mode: 0.5~40Hz
Surgical mode: 1~20Hz
Sensitivity 2V/mV ±5%
MAP Analog Signal Output
Output voltage DC 0 to 5V (0 to 500mmHg)
Output voltage error ±5%
CVP Analog Signal Output
Output voltage DC 0 to 5V (0 to 100mmHg)
Output voltage error ±5%
Alarm Limit Range Step
CCO High (Low limit + 0.1) to 25 L/min 0.1 L/min
CCO Low 0.3 to (high limit - 0.1)L/min
CCI High (Low limit + 0.1) to 15 L/min/m
2
0.1 L/min/m
2
CCI Low 0.1 to (high limit - 0.1)L/min/m
2