C - 16 BeneVision N Series Patient Monitor Operator’s Manual
C.1.13.3 Microstream CO
2
Default Settings
C.1.13.4 Mainstream CO
2
Default Settings
C.1.14 Gas Default Settings
Operating Mode Measure
Item Default Setting
Item Default Setting
BTPS Compen Off
Maximum Hold 20 sec
Auto Standby Off
Operating Mode Measure
Item Default Setting
Maximum Hold 10 sec
O
2
Compensation Off
Balance Gas Room Air
AG Compensation 0%
Operating Mode Measure
Item Default Setting
EtCO
2
Alarm switch On
High limit Adult and pediatric: 50 mmHg
Neonate: 45 mmHg
Low limit Adult and pediatric: 25mmHg
Neonate: 30mmHg
Priority Med
Alarm Outputs Off
FiCO
2
Alarm switch On
High limit 4 mmHg
Priority Med
Alarm Outputs Off
EtO
2
Alarm switch On
High limit 88%
Low limit 18%
Priority Med
Alarm Outputs Off