EasyManuals Logo

Mindray BeneVision N Series User Manual

Mindray BeneVision N Series
528 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #472 background imageLoading...
Page #472 background image
C - 8 BeneVision N Series Patient Monitor Operator’s Manual
NIBP-D Alarm switch On
High limit Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 60 mmHg
Low limit Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
Priority Med
Alarm Outputs Off
NIBP-M Alarm switch On
High limit Adult: 110 mmHg
Pediatric: 90 mmHg
Neonate: 70 mmHg
Low limit Adult: 60 mmHg
Pediatric: 50 mmHg
Neonate: 25 mmHg
Priority Med
Alarm Outputs Off
NIBP-S Extreme Alarm switch Off
High limit Adult: 175 mmHg
Pediatric: 130 mmHg
Neonate: 95 mmHg
Low limit Adult: 75 mmHg
Pediatric: 60 mmHg
Neonate: 35 mmHg
Priority High
Alarm Outputs Off
NIBP-D Extreme Alarm switch Off
High limit Adult: 105 mmHg
Pediatric: 80 mmHg
Neonate: 65 mmHg
Low limit Adult: 35 mmHg
Pediatric: 30 mmHg
Neonate: 15 mmHg
Priority High
Alarm Outputs Off
NIBP-M Extreme Alarm switch Off
High limit Adult: 125 mmHg
Pediatric: 100 mmHg
Neonate: 75 mmHg
Low limit Adult: 45 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
Priority High
Alarm Outputs Off
Item Default Setting

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals