C - 14 BeneVision N Series Patient Monitor Operator’s Manual
C.1.11 CCO Default Settings (Vigilance/Vigileo/EV1000)
C.1.12 ScvO
2
/SvO
2
Default Settings (Vigilance/Vigileo/EV1000)
Sweep 25 mm/sec
Select Parameter C.I., SQI, SVI, SVRI, TFC
Item Default Setting
Item Default Setting
CCO Alarm switch On
High limit 14.0
Low limit 2.0
Priority Med
Alarm Outputs Off
CCI Alarm switch On
High limit 10.0
Low limit 1.0
Priority Med
Alarm Outputs Off
SVR Unit DS/cm
5
Select Parameter Vigilance: CCI, SVRI, EDVI, SVI, RVEF
Vigileo: CCI, SVRI, SVV, SVI, RVEF
EV1000: CCI, GEF, SVRI, ELWI, GEDI
Item Default Setting
SvO
2
/ScvO
2
Alarm switch On
High limit 90
Low limit 40
Priority Med
Alarm Outputs Off