9-22 Structure and Assembly/Disassembly
9.3.9.6 Digital Keyboard Module (for XP)
1. As shown in the figure, unscrew one M3X8 screw to remove the fixed piece of control panel.
2. Pull out signal wire of small keyboard and unscrew six M3X8 screws to remove digital
keyboard module.