EasyManuals Logo

Mindray DC-8 Service Manual

Mindray DC-8
287 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #266 background imageLoading...
Page #266 background image
C-2 Description of Self-diagnosis Test Items
The interface can’t be closed manually after the test is started, until the test is completed. If the test
failed, please turn off the power as soon as possible.
Suggestion when test failed:
Check if PC module heat sink was correctly installed, if not, reinstalled it.
C.1.3 Z0301 PC Module and MF FPGA Interconnection Test
Test Content
Test if the LPC communication between PC module and MF FPGA is normal.
Suggestion when test failed:
1. Check if the following listed hardware and the related circuits are normal, check if the interfaces
are well welded and if the module is working normally.
2. Replacing the digital board is recommended.
Board
Connector
Digital board
U42, J1
C.1.4 Z0302 SVIDEO Buffer DDRII Test
Test Content
The super stratum test item: Z0201 PC Module and MF FPGA Interconnection Test
Test if SVIDEO DDR II works normally.
Suggestion when test failed:
1. Check if the following listed hardware and the related circuits are normal, check if the interfaces
are well welded and if the module is working normally.
2. Replacing the digital board is recommended.

Table of Contents

Other manuals for Mindray DC-8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-8 and is the answer not in the manual?

Mindray DC-8 Specifications

General IconGeneral
BrandMindray
ModelDC-8
CategoryMedical Equipment
LanguageEnglish

Related product manuals