6 - 10 Operator’s Manual
6 Image Acquisition
Priority
This function is used to set levels of the flow display, to display the grayscale signal or color signal.
The color image is preferred with higher value; while grayscale signals are displayed with the lower
value.
Line Density
The function determines the quality and information of the image.
The higher the line density is, the higher the resolution becomes.
Invert
To set the display mode of the color flow. The color scale will be inverted when the function is
turned on.
B/C Align
To set and constrain the maximum width of the B mode image to that of the Color ROI.
Packet Size
This function is an indication of the ability to detect flow, which is used to adjust the accuracy of
color flow.
The higher the sensitivity is, the more sensitive indication for low-velocity flow becomes.
Dynamic Range
This function is to adjust the transformation of echo intensity into color signal.
Increasing dynamic range will lead to higher sensitivity to low-power signals, thus enhances the
range of signals to display.
iTouch
To optimize image parameters as per the current tissue characteristics for a better image effect.
Smart Track
To optimize image parameters as per the current tissue characteristics for a better image effect.
When Smart Track is turned on, the system optimizes ROI angle and position automatically to
achieve an active tracking by reducing the impact of patient respiratory movement.
Dual Live
This function is used to display Power image synchronously.
Dual
To enter the Dual-split display mode, and to switch between the windows.
V 1:1
This function is to display images in vertical format in the dual-split mode. After the feature is
enabled, one image appears above, and the other image appears below.
In the dual-split mode, tap [V 1:1] to enable this function.
HR Flow
Enhance tiny vessel display to analyze the blood supply of the vessel in pathological organ.
Tap [HR Flow] to complete the adjustment ([HR Flow] is highlighted after it being enabled).