13 Probes and Biopsy
Operator’s Manual 13 - 17
13.1.7 Storage and Transportation
When all examinations for the day have been completed, confirm that the probe is in good
condition. After disinfecting the probe, confirm that the probe is in good condition and stored in a
suitable place
• To prevent the probe from being damaged, DO NOT store it in locations where it may be
exposed to:
– Direct sunlight or X-rays
– Sudden changes in temperature
–Dust
– Excessive vibration
– Heat generators
• When the probe is sent to MINDRAY Customer Service Department or sales representative for
repair, be sure to disinfect it and keep it in the carrying case to prevent infection.
L15-3RCs Operating 0℃~40℃ 20%~85%RH 700hPa~1060hPa
Storage and
transportation
-20℃~55℃ 20%~95%RH 700hPa~1060hPa
L16-3RCs Operating 0℃~40℃ 20%~85%RH 700hPa~1060hPa
Storage and
transportation
-20℃~55℃ 20%~95%RH 700hPa~1060hPa
L24-6RCs Operating 0℃~40℃ 20%~85%RH 700hPa~1060hPa
Storage and
transportation
-20℃~55℃ 20%~95%RH 700hPa~1060hPa
L24-6Hs Operating 0℃~40℃ 20%~85%RH 700hPa~1060hPa
Storage and
transportation
-20℃~55℃ 20%~95%RH 700hPa~1060hPa
ELC13-4s Operating 0℃~40℃ 20%~85%RH 700hPa~1060hPa
Storage and
transportation
-20℃~55℃ 20%~95%RH 700hPa~1060hPa
SP5-1Ns Operating 0℃~40℃ 20%~85%RH 700hPa~1060hPa
Storage and
transportation
-20℃~55℃ 20%~95%RH 700hPa~1060hPa
i5M Operating 0℃~35℃ 20%~85%RH 700hPa~1060hPa
Storage and
transportation
-20℃~45℃ 20%~85%RH 700hPa~1060hPa
Table 13-2 Probe Environmental Conditions
Probe
Model
Conditions Ambient
temperature
Relative humidity
(no condensation)
Atmospheric
pressure