1 Important Information
Operator’s Manual 1 - 7
• Please use the disinfection or sterilization solution recommended in this
operator's manual; otherwise Mindray will not be liable for damage caused
by other solutions. If you have any questions, please contact Mindray
Customer Service Department or sales representative.
• Do not use pre-lubricated condoms as a sheath. Lubricant may not be
compatible with the probe material and damage may result.
• The damage of the transducer may be caused by the contact of improper
gel or cleaner:
– DO NOT dip the transducer in the strong polar solution of ethanol,
chloride of lime, ammonium chloride, acetone and formaldehyde.
– DO NOT contact the transducer with solution or ultrasound gel
containing oily medium such as mineral oil or lanoline.
• Read the following precautions to prevent the probe from malfunction:
– Before connecting or disconnecting the probe, freeze or turn off the system.
– Clean and disinfect the probe before and after each examination.
– After the examination, wipe off the ultrasound gel thoroughly. Otherwise, the ultrasound
gel may solidify and the image quality would be degraded.
• Repeated disinfection will eventually damage the probe, please check the probe performance
periodically.
1.2 Latex Alert
When choosing a probe sheath, it is recommended that you directly contact CIVCO for obtaining
information regarding probe sheaths, pricing, samples and local distribution.
For CIVCO information, please contact the following:
CIVCO Medical Instruments
Tel: 1-800-445-6741
www.civco.com
Allergic reactions in patients sensitive to latex (natural rubber) may range from
mild skin reactions (irritation) to fatal anaphylactic shock, and may include
difficulty breathing (wheezing), dizziness, shock, swelling of the face, hives,
sneezing, or itching of the eyes (FDA Medical Alert on latex products, “Allergic
Reactions to Latex-containing Medical Devices”, issued on March 29, 1991).
1.3 Parts that can be used within patient
environment
• The ultrasound system