13 Probes and Biopsy
Operator’s Manual 13 - 35
a. Put on the sterile probe sheath.
b. Select the needle-guided bracket, and match the groove with the tab of the probe. Mount
the bracket onto the probe.
c. Rotate the retaining nut to fix the bracket and the probe.
2. Adjust the needle angle to the proper shift as required:
a. Loosen the angle pinch nut.
b. Adjust the angle block to the desired level.
c. Tighten the angle pinch nut.
3. Insert the biopsy needle:
a. Hold the probe. Press the biopsy needle to separate needle guided V-shaped block from
pressure position of the needle
b. Put the needle into the needle guided-bracket, and the needle leans to V-shaped block.