A-10
A.7.5 NIBP
Standards
Meet standards of EN60601-2-30/IEC60601-2-30, EN1060-1, EN1060-3,EN1060-4 and
SP10
Technique Oscillometry
Mode of operation Manual, Auto and STAT
Auto mode repetition intervals
1 min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15 min, 20 min, 30 min, 1 h, 1.5 h, 2 h, 3 h, 4
h, 8 h
STAT mode cycle time 5 min
Max measurement time 90 s
Measurement ranges
(mmHg)
Big Medium Small
Systolic: 25 to 285 25 to 230 25 to 135
Diastolic: 10 to 250 10 to 200 15 to125
Mean: 15 to 260 15 to 215 10 to 115
Accuracy
Max mean error: ±5 mmHg
Max standard deviation: 8 mmHg
Resolution 1mmHg
Initial cuff inflation pressure range
(mmHg)
Big
Medium:
Small:
80 to 280
80 to 210
60 to 140
Default initial cuff inflation pressure
(mmHg)
Big
Medium:
Small:
160
140
90
Software overpressure protection
Big
Medium:
Small:
297±3 mmHg
240±3 mmHg
147±3 mmHg
PR
Measurement range 40 to 240 bpm
Resolution 1 bpm
Accuracy ±3bpm or ±3%, whichever is greater
Alarm limit Range (mmHg) Step (mmHg)
Sys High
Big: (low limit+5) to 270
Medium: (low limit+5) to 200
Small: (low limit+5) to 135
5
Sys Low 40 to (high limit-5)
Mean High
Big: (low limit+5) to 230
Medium: (low limit+5) to 165
Small: (low limit+5) to 110
Mean Low 20 to (high limit-5)
Dia High
Big: (low limit+5) to 210
Medium: (low limit+5) to 150
Small: (low limit+5) to 100
Dia Low 10 to (high limit-5)