A-11
*Measurement accuracy verification: the blood pressure measurements measured with this device are in compliance with
the American National Standard for Electronic or Automated Sphymomanometers (ANSI/AAMI SP10) in terms of mean error
and stardard deviation by comparing with intra-arterial measurements (depending on the configuration) in a typical patient
population.
A.7.6 Temp
Standards Meet standard of EN12470-4
Technique Thermal resistance
Measurement range 0 to 50 ℃ (32 to 122 ℉)
Resolution 0.1 ℃
Accuracy ±0.1 ℃(without probe)
Refreshing rate 1 s
Minimum time for accurate
measurement
Body surface: <100 s
Body cavity: <80 s
Alarm limit Range Step
T1/T2 High (low limit +1) to 50 ℃
0.1 ℃
T1/T2 Low 0.1 to (high limit -1) ℃
TD High 0.1 to 50 ℃
A.7.7 IBP
Standards Meet standard of EN60601-2-34/IEC60601-2-34.
Technique Direct invasive measurement
IBP
Measurement range -50 to 300 mmHg
Resolution 1 mmHg
Accuracy ±2% or ±1 mmHg, whichever is greater (without sensor)
Refreshing rate 1 s
Pressure transducer
Excitement voltage 5 VDC, ±2%
Sensitivity 5 μV/V/mmHg
Impedance range 300 to 3000Ω
Alarm limit Range (mmHg) Step (mmHg)
Sys High
(low limit + 2) to 300
1
Mean High
Dia High
Sys Low
-50 to (high limit – 2) Mean Low
Dia Low