5-13
To remove Mindray CO2 module:
1. Unscrew the two PT2×6 screws. Release the plastic clamp and take out the tubings.
2. Take out the watertrap assembly and move it through the fixer in the way as shown
below.
3. Unplug the communication cable from the multi-parameter board. Unplug the tubing
from the gas nipple. Then unscrew the two nuts and pull out the Mindray CO2
assembly.