16 - 36 Operator’s Manual
16 Probes and Biopsy
b. Put the needle into the needle guided-bracket, and the needle leans to V-shaped block.
c. Hold the probe, and release the pressure position of the needle. Adjust the needle-type
adjusting nut manually (following the direction of the arrow). The needle moves smoothly
at the vertical direction due to its gravity.
4. Release the needle from the bracket: