ECG-Arrhythmia Troubleshooting ECG – Arrhythmia
9 - 38 V Series Operating Instructions
Base Line Wander Patient moving excessively. Secure lead wires and cable to
patient.
Patient's respiration. Reposition electrodes.
Electrodes dry or loose. Re-prep skin and apply fresh, moist
electrodes.
Static build up around patient. Check with local biomedical
personnel.
ECG Filter set to ST or Diagnostic
mode.
Set ECG Filter to “Monitor” mode.
MESSAGE/PROBLEM REASON SOLUTION