EasyManuals Logo

Mindray V21 User Manual

Mindray V21
587 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #520 background imageLoading...
Page #520 background image
V Series Operating Instructions 29 - 55
Product Specifications Electromagnetic Capability
29.19 Electromagnetic Capability
29.19.1 EMC Safety
The V 12/V 21 monitor needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information in this Instructions For Use. Portable and mobile RF
communications equipment can affect the V 12/V 21 monitor.
The V 12/V 21 is suitable for use in all establishments other than domestic establishments and those
directly connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
Radiated Emission
Complies with the requirements of CISPR 11 (EN 55011:2007) Group 1, Class A.
Immunity Against Radiated Electromagnetic Fields
Complies with the requirements of IEC 61000-4-3:2001: 80 MHz to 2.5 GHz, 3V/m, 80% AM @ 2 Hz or
modulation frequency significant for the equipment under test.
29.19.2 V 12/V 21
The V 12/V 21 meets the requirements of IEC 60601-1-2/EN 60601-1-2.
NOTE: The V 12/V 21 needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC information
provided below.
NOTE: Portable and mobile RF communications equipment can affect the V
12/V 21. See tables 29-33 through 29-36 that follow.
TABLE 29-33
GUIDANCE AND DECLARATION - ELECTROMAGNETIC EMISSIONS
The V 12/V 21 is intended for use in the electromagnetic environment specified below. The customer or the
user of the V 12/V 21 should assure that it is used in such an environment.
EMISSIONS
TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
RF emissions
CISPR 11
Group 1 The V 12/V 21 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A The V 12/V 21 is suitable for use in all establishments other
than domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes
Harmonic
emissions IEC
61000-3-2
Class A
Voltage
fluctuations/
Flicker
emissions IEC
61000-3-3
Complies

Table of Contents

Other manuals for Mindray V21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V21 and is the answer not in the manual?

Mindray V21 Specifications

General IconGeneral
BrandMindray
ModelV21
CategoryMedical Equipment
LanguageEnglish

Related product manuals