Alarm Limits Alarms
8 - 10 V Series Operating Instructions
8.2.2.7 Neuro Tab
TABLE 8-7 lists the alarm limit ranges for the BIS parameter, which are contained in the Neuro tab.
8.2.3 Auto-Set All Alarms Button
The Auto-Set All button automatically recalculates high and low alarm limit settings to the percent
indicated in TABLE 8-8 above and below the patients current measured value for the selected
parameter.
NOTE: The systolic, diastolic, and mean NIBP values flash in the corresponding
alarm priority color when an alarm is active.
8.2.4 Restore All Defaults Button
Restores all alarm limits to the departmental default values.
TABLE 8-7
PARAMETER
ALARM
LIMIT ADULT PEDIATRIC NEONATE
BIS
Invalid data: “- -”
High Limit Off, 10 to 99
Default: Off
Off, 10 to 99
Default: Off
N/A
Low Limit Off, 1 to 90
Default: Off
Off, 1 to 90
Default: Off
N/A
TABLE 8-8
ALARM
PERCENT INDICATOR
(APPROXIMATE)
Heart Rate 30%
Respiration 30%
IBP Systolic 20%
IBP Diastolic 20%
IBP Mean 20%
NIBP Systolic 20%
NIBP Diastolic 20%
NIBP Mean 20%
ET CO
2
(mmHg) 20%
ET CO
2
(%) 20%
INSP CO
2
(mmHg) 20%
INSP CO
2
(%) 20%
ET CO
2
(kPa) 20%
INSP CO
2
(kPa) 20%
Temp (°F) 3%
Temp (°C) 3%
Delta Temp (°F) 25%
Delta Temp (°C) 25%

Table of Contents