V Series Operating Instructions 8 - 9
Alarms Alarm Limits
8.2.2.5 CCO/Hemo Tab
Refer to the Addendum to V Series Operating Instructions for additional information.
8.2.2.6 Oximetry
TABLE 8-6 lists the available alarm limits for the SpO
2
parameter in the Oximetry tab.
NOTE: For the rSO2 alarm limit ranges, refer to the Addendum to V Series
Operating Instructions for additional information.
Sev ET
Unit of Measure: %
Resolution: 0.1%
High Limit Off, 0.2 to 8.0
Default: Off
Off, 0.2 to 8.0
Default: Off
Off, 0.2 to 8.0
Default: Off
Low Limit Off, 0.0 to 7.8
Default: Off
Off, 0.0 to 7.8
Default: Off
Off, 0.0 to 7.8
Default: Off
O2 Insp
Unit of Measure: %
Resolution: 1%
High Limit Off, 40 to 100
Default: Off
Off, 40 to 100
Default: Off
Off, 40 to 100
Default: Off
Low Limit Off, 18 to 60
Default: 18
Off, 18 to 60
Default: 18
Off, 18 to 60
Default: 18
O2 ET
Unit of Measure: %
Resolution: 1%
High Limit Off, 40 to 100
Default: Off
Off, 40 to 100
Default: Off
Off, 40 to 100
Default: Off
Low Limit Off, 10 to 60
Default: Off
Off, 10 to 60
Default: Off
Off, 10 to 60
Default: Off
N2O Insp
Unit of Measure: %
Resolution: 1%
High Limit Off, 10 to 80
Default: 80
Off, 10 to 80
Default: 80
Off, 10 to 80
Default: 80
Low Limit Off, 5 to 70
Default: Off
Off, 5 to 70
Default: Off
Off, 5 to 70
Default: Off
N2O ET
Unit of Measure: %
Resolution: 1%
High Limit Off, 10 to 80
Default: Off
Off, 10 to 80
Default: Off
Off, 10 to 80
Default: Off
Low Limit Off, 5 to 70
Default: Off
Off, 5 to 70
Default: Off
Off, 5 to 70
Default: Off
TABLE 8-5
PARAMETER
ALARM
LIMIT ADULT PEDIATRIC NEONATE
TABLE 8-6
PARAMETER
ALARM
LIMIT ADULT PEDIATRIC NEONATE
SpO2
Unit of Measure: %
Resolution: 1%
Invalid data: “- -”
High Limit Off, 75 to 101
Default: 101
Off, 75 to 101
Default: 101
Off, 75 to 101
Default: 101
Low Limit 70 to 99
Default: 88
70 to 99
Default: 90
70 to 99
Default: 92